6. While the samples are incubating, perform a 30 sec quick-spin

centrifugation of the detection antibody (10x for multiplex assays,
100x for singleplex assays) prior to pipetting to collect the entire
required volume at the bottom of the vial.

7. Prepare a sufficient volume of detection antibodies (1x) using

detection antibody diluent. Refer to the example tables below.

NOTE: The detection antibody stock concentration is 10x in the
multiplex assays and 100x in the singleplex assays. Always check
the detection antibody concentration on the vial label before diluting.

8. After incubating the samples, slowly remove and discard the sealing

tape, then

vacuum-filter.

9.

Wash 3 times.

10. Vortex the detection antibodies gently and add 25 µl to each well.

Incubate for 30 min.

Example Detection Antibody Calculations (10x stock)

# of Wells

10x Detection

Antibody (µl)

Total Volume (µl)

96

48

32

24

300

150

100

75

2,700

1,350

900

675

3,000

1,500

1,000

750

Detection Antibody

Diluent (µl)

Example Detection Antibody Calculations (100x stock)

# of Wells

100x Detection

Antibody (µl)

Total Volume (µl)

96

48

32

24

30

15

10

7.5

2,970

1,485

990

742.5

3,000

1,500

1,000

750

Detection Antibody

Diluent (µl)