Section 8

Control Preparation

Two vials of lyophilized acute phase controls are provided in each

multiplex Bio-Plex Pro™ acute phase assay kit. This section provides

instructions on how to reconstitute the Iyophilized control. The product

insert provided with the assay lists the concentration ranges of the

prepared controls.

Prepare Acute Phase Controls
To ensure optimal assay performance, the acute phase controls should be

prepared in a manner consistent with that used to prepare the acute phase

standards.

1. Gently tap the glass vial containing the lyophilized acute phase

control on a solid surface to ensure the pellet is at the bottom.

2. Refer to the steps below for reconstituting the 5-plex and 4-plex

controls. Use the standard/control diluent prepared in the previous

section (Section 7, step 2). Do not use assay buffer to reconstitute

the controls.

5-plex control:

Reconstitute one vial of lyophilized control with

1 ml of 5-plex standard/control diluent. This solution will be used

directly in the assay; no additional dilutions are required.

4-plex control:

Reconstitute one vial of lyophilized control with

2 ml of 4-plex standard/control diluent. This solution will be used

directly in the assay; no additional dilutions are required.

3. Gently vortex 1-3 sec and incubate on ice for 60 min. For optimal

assay performance, be consistent with the incubation time.

4. Expected concentration ranges of the prepared controls are provided

in the standard and control values product insert.