Fig. 1. Bio-Plex sandwich immunoassay.

Data Acquisition and Analysis
Data from the reactions are acquired using a Bio-Plex system or similar
Luminex-based reader. When a multiplex assay suspension is drawn into
the Bio-Plex 200 reader, for example, a red (635 nm) laser illuminates the
fluorescent dyes within each bead to provide bead classification and thus
assay identification. At the same time, a green (532 nm) laser excites PE
to generate a reporter signal, which is detected by a photomultiplier tube
(PMT). A high-speed digital processor manages data output, and
Bio-Plex Manager

software presents data as median fluorescence

intensity (MFI) as well as concentration (pg/ml). The concentration of
analyte bound to each bead is proportional to the MFI of reporter signal.

Using Bio-Plex Data Pro

software, data from multiple instrument runs

can be combined into a single project for easy data management, quick
visualization of results, and simple statistical analysis.

Biomarker

of Interest

Magnetic Bead

Capture

Antibody

Biotinylated

Detection

Antibody

Phycoerythrin

Fluorescent

Reporter

Streptavidin