beautypg.com

Diluent – Bio-Rad Rat Diabetes Assays User Manual

Page 47

background image

45

2. Determine the volume of 1x diabetes and cytokine detection antibodies needed.

a) Each well requires 25 μl of detection antibodies (1x): _______ x 25 μl = ______ μl

1 13

b) Include 25% excess to ensure enough volume: _______ μl x 0.25 = ______ μl

13 14

c) Total volume of 1x detection antibodies: ______ μl + ______ μl = ______ μl

13 14 15

d) Volume of 20x detection antibodies required from each diabetes tube(s):

______ μl / 20 = ______ μl

15 16

e) Volume of 10x detection antibodies required from each cytokines tube(s):

______ μl / 10 = ______ μl

15 17

f) Total volume of diabetes detection antibodies stock required: ____ μl x ____ =_____ μl

16

2 18

g) Total volume of cytokine detection antibodies stock required: ____ μl x ____ =_____ μl

17 3 19

h) Total volume of combined detection antibodies required:

_____ μl + ______ = _____ μl

18

19 20

i) Volume of detection antibody

diluent

required: ______ μl – _____ μl = _____ μl

15 20 21

3. Determine the volume of 1x streptavidin-PE needed.

a) Each well requires 50 μl of streptavidin-PE (1x): _________ x 50 μl = _________ μl

1 15

b) Include 25% excess to ensure enough volume: _______ μl x 0.25 = ______ μl

15 16

c) Total volume of 1x streptavidin-PE: ______ μl + ______ μl = ______ μl

15 16 17

d) Volume of 100x streptavidin-PE required: _______ μl / 100 = ______ μl

17 18

e) Volume of assay buffer required: ______ μl – ______ μl = __________ μl

17 18

19

This manual is related to the following products: