lmportant Considerations

Assay Procedures

n

Please pay close attention to vortexing, shaking, and incubation

times and to Bio-Plex

®

reader PMT (RP1) setting, as these have been

optimized specifically for the cell signaling assays

n

For optimal performance, use only reagents specific for Bio-Plex Pro

cell signaling assays. Reagents in other Bio-Plex assay panels have not
been validated for use in the cell signaling assays

n

Do not reuse diluted (1x) coupled beads, detection antibodies, or

streptavidin-PE

n

Wash as outlined in Table 5. Incomplete washes may cause

assay variation

n

If the data are not acquired immediately, the assay plate

may be stored

at 4°C for up to 24 hr protected from light

Assay Quick Guide
Each assay kit comes complete with a printed Bio-Plex Pro Assay Quick Guide
(bulletin 10024930), which can be used to prepare and run a full 1 x 96-well
assay plate. Users can also download a copy at www.bio-rad.com/bio-plex.

Bead Regions and Multiplexing Compatibility

n

Bead regions for all analytes are listed in Table 12

n

Compatible singleplex assays may be mixed to create a multiplex assay

n

Do not mix phosphoprotein assays with corresponding total target

assays (for example, phospho-Akt and total Akt)

Instruments and Software
The Bio-Plex Pro cell signaling assays described in this manual are
compatible with all currently available Luminex-based life science
research instruments. Assays can be read and analyzed with either
Bio-Plex Manager

software or Luminex xPONENT software (Section 5,

Read the Plate, under Detailed Instructions).