beautypg.com

Bio-Rad Bio-Plex Pro™ Magnetic Cell Signaling Assays User Manual

Page 32

background image

30

Table 13. Troubleshooting guide (continued).

Problem

Possible Causes

Possible Solutions

High background

Prolonged incubation of

Follow the procedure incubation

detection antibodies and/or

time precisely

streptavidin-PE

Wash steps performed

Perform washes as described in

incorrectly or insufficient

the assay instructions

washing volume

High assay CV

Bottom of filter plate not dry

Dry the bottom of the filter plate

with absorbent paper towel

(preferably lint-free) to prevent

cross-well contamination

Contamination with wash

Be careful not to splash wash buffer

buffer during wash steps

from well to well. Filter wells
completely to remove residual
buffer if using filter plate. Reduce
microplate shaking speed to
minimize splashing

Cell debris in lysate not cleared Centrifuge at 15,000 x g for 10 min

at 4°C to remove cellular debris

Shaking speed too high during Fully resuspend bead mixture at

assay incubation

900–1,100 rpm for 30 sec, then turn
down to 300–450 rpm for the
specified incubation time

Reagents and assay

Bring all reagents and assay

components not equilibrated

components to room temperature

to room temperature

prior to dispensing