Troubleshooting guide – Bio-Rad Bio-Plex Pro™ Magnetic Cell Signaling Assays User Manual
Page 30
28
Troubleshooting Guide
This troubleshooting guide addresses problems that may be encountered
with Bio-Plex Pro
™
assays. If you experience any of the problems listed
below, review the possible causes and solutions provided. Suboptimal
assay performance may also be due to the Bio-Plex
®
suspension array
reader. To eliminate this possibility, use the Bio-Plex validation kit to assist
in determining if the array reader is functioning properly.
Table 13. Troubleshooting guide.
Problem
Possible Causes
Possible Solutions
Low signals for
Sample protein concentration
Verify sample protein concentration
experimental
too low or too high
is within assay range. Optimization
samples
of protein concentration may be
needed based on targeted protein
expression level
Poor cell lysate quality
Prepare fresh lysate accordingly
Low signals for
Incorrect dilution of detection
Check the calculations and be
experimental samples antibody or streptavidin-PE
careful to add the correct volumes
and Bio-Rad cell
lysate controls
Expired Bio-Plex reagents
Check that reagents have not
were used
expired. Use new or non-expired
components
Incorrect incubation
Incubations should be at room
temperature
temperature (20–22°C)
Insufficient incubation time
Adhere to the recommended
incubation time
Beads lost
Use recommended magnetic
washer settings with correct
magnet soaking time
continues