Bio-Rad PowerPac™ Universal Power Supply User Manual
Page 23

Step Procedure
Description
4.
Edit a step
a.
Set the Constant mode: voltage (V),
current (A) or power (W).
b.
Enter the constant and non-constant
values in the Volts, mAmps and Watts
fields.
c.
Select whether the run will be timed
(Untimed box is not checked) or
un-timed (the Untimed box is checked).
d.
Change the method name as needed.
Check the Rename box to rename the
method or leave it un-checked to save
the edited method with the new name.
e.
Select another step to edit from the
"Step" drop-down list and edit the step.
f.
Press Save to save the method.
Other Buttons
Delete: Deletes the entire method.
Delete Step: Deletes the current step.
Methods: Returns to Method List screen.
5.
Transmit the method
Press Beam to transmit the method to the
PowerPac Universal. Note that if nine
methods already exist on the PowerPac
Universal that the new method will be
transmitted to flash memory so it can be run,
but it will not be saved to the method list.
4.2.2 Receive and Transmit Data
Step Procedure
Description
1.
Start the software
Start the PowerPac Remote software and
press Run Data
17
Method: SDS-PAGE
Step # 1
Const Mode
Rename
Volts: 100
mAmps: 2500
Watts: 500
Vhours: 00000
Untimed
Save
Delete
Beam
Methods
Delete Step
V
A
Timing
T
Vh
W
Bio-Rad Laboratories
PowerPac Remote Control
Select/Edit Methods...
Run Data