Bio-Rad PowerPac™ Universal Power Supply User Manual
Page 15

Step Procedure
Description
2.
Select the programming
On the Home screen, press the METHODS
mode
soft key to display the Method List screen.
3.
Create a new method
Press the NEW soft key on the Methods List
screen to create a new method. Note that if
there are already nine methods defined in the
list, an existing method will need to be deleted
before a new method can be added to the list.
To delete a method use the up and down
arrow keys to select a method (identified by *)
and then press the DELETE soft key. Press
the OPEN soft key to open the selected,
stored method and display the method.
4.
Name the new run
Press the up arrow key to select the method
name field. Note that the soft keys have now
changed to CANCEL and OK. Use the CE
button to delete the default name and the
keypad to enter a method name. Refer to
Section 2 for help using the keypad. Press the
CANCEL soft key to abort naming the method
or the OK soft key to accept the new name.
Note that OK accepts the name but is not
stored into memory until the method is saved.
5.
Select a step to edit
Press the EDIT key to open the selected
empty step (identified by *).
6.
Set the constant
Use the soft keys to select the constant
parameter for the
parameter and the keypad to enter its value. If
current step
the method is complete go to Step 10,
otherwise, press the EDIT key. Note that the
maximum default values for the non-constant
parameters are listed on the right side of the
screen and the time mode at the bottom.
ConstV:
Used to run with constant voltage.
ConstA:
Used to run with constant current.
ConstW: Used to run with constant power.
9
SELECT MODE
MANUAL
METHODS
METHOD LIST
...EMPTY 1...
...EMPTY 2...
...EMPTY 3...
OPEN
DELETE
NEW
METHOD: UNTITLED
S1 (EMPTY)
*
METHODS
EDIT NAME: NAME_
S1 (EMPTY)
CANCEL
OK
METHOD: NAME
S1 (EMPTY)
*
METHODS
DELETE
SAVE
TIME: UNTIMED
S1/1 EDIT CONSTANT
CONSTA CONSTW
2500
500
mA
W
CONST
V
0 V