100 v – Bio-Rad PowerPac™ Universal Power Supply User Manual
Page 16

Step Procedure
Description
7.
Enter non-constant
Use the soft keys to select a non-constant
parameter limits for
parameter and the keypad to enter its value.
the current step
If the method is complete and will be run
un-timed go to Step 10, otherwise, press the
EDIT key. Note: to change the constant value,
press its soft key and enter a new value.
ConstV:
Used to change the voltage
constant value.
LimitA:
Used to change the maximum
default value for current limit.
LimitW:
Used to change the maximum
default value for power limit.
8.
Select the time mode
Use the soft keys to select a time mode and
the keypad to enter the run time. Press EDIT.
HRS:
Run time is in units of hours.
VHOURS: Run time is in units of volt-hours.
UNTIMED: Run is continuous. Note that steps
programmed after an un-timed
step are ignored.
9.
Add a new step or save
To program additional steps press the down
the method
arrow key to select the empty step and repeat
steps 5–8.
If the method will not be saved, go directly to
Step 10, otherwise press the SAVE soft key
and then go to Step 10. Press the DELETE
soft key to delete the selected step. Press the
METHODS soft key to return to the Method
List screen.
10.
Start the run
Press
to start a run.
11.
Monitor the run
The run screen is used to monitor run
parameters or to pause a run. The current
step versus total steps programmed, voltage,
current and power values, elapsed time versus
the programmed time, and total elapsed time
of the method are displayed on the run screen.
Press the PAUSE soft key to edit the run
parameters (see Section 3.3.3). Press
to
terminate the run. Note that any edits
made during the run will not be saved unless
SAVE is pressed at the end of the run.
12.
View the run
The "Run Completed" or "Run Terminated"
information
screen is displayed at the end of a run. The
date and time, the final running parameter
values, the total elapsed time in hours and
volt-hours are displayed on the screen. Press
the RERUN soft key to repeat the run or the
EXIT soft key to display the Method Step List.
After the run is finished, the run data may be
transmitted to a PDA or PC (see Section 4.2.2).
10
METHOD: NAME
S1 100V 00:30 HRS
S2 (EMPTY)
*
METHODS
DELETE
SAVE
00:00/00:30 00:00 HRS
PAUSE
53
5
mA
W
100 V
RUN
: S1/2
RUN COMPLETED AT:
05/27/03 12:34:01
200V 106mA 21W
Time: 00:01 Vh: 00003
RERUN
EXIT
TIME: UNTIMED
S1/1 EDIT LIMITS
LIMITW
2500
500
mA
W
CONSTV
100 V
LIMIT
A
TIME: UNTIMED
S1/1 EDIT TIME
UNTIMED
2500
500
mA
W
HRS
100 V
VHOURS