100 v – Bio-Rad PowerPac™ Universal Power Supply User Manual
Page 18
Step
Procedure
Description
6.
Set the non-constant
Use the soft keys to select a non-constant
parameter limits for
parameter and the keypad to enter its value.
the current step
If the method is complete skip to Step 9,
otherwise, press the EDIT key. Note: to
change the constant value, press its soft key
and enter a new value.
ConstV:
Used to change the voltage
constant value.
LimitA:
Used to change the current limit
value.
LimitW:
Used to change the power limit
value.
7.
Select the time mode
Use the soft keys to select a time mode and
the keypad to enter the run time. Press the
EDIT key.
HRS:
Run time is in units of hours.
VHOURS: Run time is in units of volt-hours.
UNTIMED: Run is continuous. Note that steps
programmed after an un-timed
step are ignored.
8.
Add a new step or
If more steps are to be added to the method,
save the method
use the down arrow key to move to the empty
step, press the EDIT key and repeat steps 5–8
to add a new step. If the method will not be
saved, go to Step 9, otherwise press the
SAVE soft key and then go to Step 9.
9.
Start the run
Press
to start a run.
10.
Monitor the run
The run screen is used to monitor run
parameters or to pause a run. The current
step versus total steps programmed, voltage,
current and power values, elapsed time versus
the programmed time, and total elapsed time
of the method are displayed on the run screen.
Press the PAUSE soft key to edit the run
parameters (see Section 3.3.3). Press
to
terminate the run. Note that any edits
made during the run will not be saved unless
SAVE is pressed at the end of the run.
12
METHOD: SDS-PAGE
S1 100V 00:30 HRS
S2 300V 00:30 HRS
S3 (EMPTY)
*
METHODS
DELETE
SAVE
00:00/00:30 00:00 HRS
PAUSE
53
5
mA
W
100 V
RUN
: S1/2
TIME: 00:30 HRS
S1/2 EDIT TIME
UNTIMED
2500
500
mA
W
HRS
100 V
VHOURS
TIME: 00:30 HRS
S1/2 EDIT LIMITS
LIMITW
2500
500
mA
W
CONSTV
100 V
LIMIT
A