Gossen GL SpectroSoft User Manual
Page 15

GOSSEN Foto- und Lichtmesstechnik GmbH
15
4
Basic Function Tool Bar
The different functions are accessible depend on the version / license of the program.
1. 2. 3. 4 . 5. 6. 7.
8. 9. 10. 11. 12. 13. 14. 15.
1. Open: previously saved measurements can be open,
2. Save group: all currently open / active measurements can be saved in one file
3. Zoom max optimizes the view of the whole spectrum chart in the
current window size.
a) Using the mouse wheel you can zoom image spectrum +/-
b) Hold Y and move the wheel to change the view in the Y-axis
c) Hold X and move the wheel to change the view of the X-axis
d) Click twice on the chart box and the view options set to max zoom
4. Measure To make measurements, or simply press the space bar. Additionally, the
GL SPECTIS 1.0 is equipped with a trigger button located on the side of the body.
5. Continuous sets the measurement procedure in continuous mode according
to preset parameters, integration time and number of
repetitions. This allows for continuous monitoring of changes in the spectrum under
certain conditions. To stop click Continuous again.
6. Serial measurement interval. Allows you to performa series of measurements with a
programmed time interval (see serial measurements chapter).
7. Serial measurement count. Allows you to performa programmed count of
measurements (see serial measurements chapter).
8. Config button allows access to set the basic parameters of measurements such as:
a) integration time;
b) number of repetitions of the measurement
c) data save path
d) range of spectral measurement
9. Indication of the temperature.
10. Indication of the signal level.