3 connection of a balance – Metrohm 787 KF Titrino User Manual
Page 59
4.1 Setting up and connecting the instruments
787 KF Titrino
55
4.1.3 Connection of a balance
The following balances can be connected to the RS232 output of the
Titrino:
Balance Cable
Sartorius MP8, MC1
6.2125.070
Shimadzu BX, BW
6.2125.080
Settings on Titrino: balance SARTORIUS
Balance: delimiter CR+LF
Ohaus Voyager, Explorer,
Analytical Plus
from Ohaus: AS017-09 (Ohaus parts number)+
6.2125.170
Settings on Titrino: balance SARTORIUS
Balance: SET BALANCE
INTERFACE
BAUD RATE 9600
DATA BITS 8
PARITY none
STOP BITS 1
Mettler AB, AG (LC-RS25)
in the scope of delivery of the balance
Mettler AM, PM
6.2146.020
additionally from Mettler:
ME 47473 Adapter and ME 42500 hand switch or ME
46278 foot switch
Mettler interface 016
Cable in scope of delivery of interface 016: red lead
to pin 3, white lead to pin 7 of the 25-pin connector
25-pole/9-pole adapter
Mettler interface 011 or 012
6.2125.020
Mettler AT
6.2146.020
Mettler PG, AB-S
6.2134.120 + 6.2125.170
Mettler AX, MX ,UMX
6.2134.120 + 6.2125.170
AND Models ER-60, 120, 180, 182
Models FR-200, 300
Models FX-200, 300, 320
with RS232 interface (OP-03)
6.2125.020
Precisa, balances with RS232C-
interface
6.2125.080
The balance type must be preselected at the Titrino with the
Balance and printer can be connected at the same time with the aid
of the 6.2125.030 adapter. The balance must then be plugged into
the "data in" receptable of the adapter.