2.8. Result calculations
787 KF Titrino
32
divisor 0.1
cond.
blank 0.0 ml
cond.
drift corr.: auto
cond.
drift value 0.0 µl/min
cond.
Divisor (
±
1000000)
in the denominator of the calculation formula. Used
mainly for the density, see page 33.
Request only in the "KFT" mode.
Blank (0.0...99.9991 mL)
The blank value is written in automatically from the
blank determination mode. Values that are corrected
by a blank value are flagged with "-b".
Request only in the "KFT" mode.
Drift correction for the result (auto, man., OFF)
With the drift correction switched on, the value
(drift*titration time) is subtracted from the KFR volume.
Values that has been corrected by the drift are flagged
with "-d".
auto: The drift value at the start of the titration is
subtracted.
man: A fixed value of the drift is entered and
subtracted.
If the drift correction is set to "man.", the drift value is
requested:
Drift value for the manual drift correction (0...99.9
µL/min)
>statistics
mean n= 20
cond.
res.tab: original
cond.
delete n= 1
cond.
Statistics calculation
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.