Metrohm 787 KF Titrino User Manual
Page 42
2.10. Data output
787 KF Titrino
38
Example of reports:
Full result report
- date, time, current run number.
- sample size, identification.
- The blank is printed out only if the value is ≠ 0 mL.
- The drift and the (-d)time are printed out only it the drift
correction has been set to "OFF".
- result
- statistics
Short result report
- date, time, current run number.
- result
- mean value with 5 single results
- standard deviation
- relative standard deviation
Additional possibilities for report outputs
In addition to the reports which are printed at the end of the titration, various other reports
can be put out. There are 2 possibilities to select the reports:
key X is the key under which the appropriate data
are entered.
List of reports:
Report
Configuration report
CONFIG
Parameter report
PARAM
Calculation data
CALC DATA
Table of the individual results for statistics
calculation
1
Result display without a printer
Results can also be viewed in the display. After a determination the calculated result and
the single results of the statistics calculations can be displayed with
date 2002-03-27 time 08:33:42 4
smpl size 1.0 g ident. 08-4
KFR volume 2.278 ml
blank 0.0315 ml
titer 5.0 mg/ml
drift auto 1.9 µl/min
(-d)time 0:43
water 1.14 %
mean(4) 1.13 %
+/-s 0.005 %
s(rel) 0.44 %
============
date 2002-03-27 time 08:35:42 5
smpl size 1.0 g ident. 08-5
(-d)time 0:45
water 1.14 %
mean(5) 1.13 %
+/-s 0.005 %
s(rel) 0.44 %
============