Page 20

2.3. Tutorial
787 KF Titrino
16
It is now necessary to match the factor to the new result unit. A table with a selection of
matched quantities for factor, divisor and unit of the sample size can be found in the
quick reference.
4 ×
<1>
2 ×
• Enter a new factor: Press
the inquiries of the "calculation" group with
display:
• Enter "1".
• As divisor enter the density of the sample.
• Exit the inquiry by pressing
sult is recalculated and outputted.
2.3.4 Drift
The drift is measured in µL/min. It specifies how much KF reagent is consumed per unit
time to keep the titration cell dry. The reagent is needed,
• to titrate moisture in the titration cell.
• to compensate the iodine consumption of slow side reactions.
The drift value is generally of the order of a few µL/min for "dry conditioned" titration cells.
Checking the condition of the solvent
The drift can be displayed during the conditioning of the titration cell with the
key:
• Press
You can follow the progress of the drift. In fresh solvent
the drift value should drop to less than 10 µL/min after
a few minutes.
• Pressing
display off.
KFT volume
3.459 ml
factor 0.1
KFT volume
3.459 ml
divisor 1
Drift 4.3
µl/min