5 Carrying out sample digestion

909 UV Digester

■■■■■■■■

29

Note

Leave the PTFE stoppers on the sample vessels until you are ready
to continue using the samples.

3 Switch off the 909 UV Digester or carry out additional sample diges-

tion processes.

Canceling sample digestion

Proceed as follows if you would like to stop a sample digestion process
before the radiating time has elapsed:

1 Press the [START / STOP] key.

The UV lamp is switched off.

2 Wait for the samples to cool (the fan switches off after 10 min).

3 Remove the sample holder with the samples from the wet end.

4 Switch off the 909 UV Digester or carry out additional sample diges-

tion processes.

Note

Opened samples smell like ammonia or fish (amine)

If an open sample smells like ammonia or fish, you need to increase the
amount of acid and/or H

2

O

2

being added.