5 Carrying out sample digestion

909 UV Digester

■■■■■■■■

27

Note

Use the provided sample vessels

Only use the sample vessels made of quartz glass (6.2414.000)
that have been provided. Normal reagent glasses do not let UV
radiation through and do not ensure complete digestion. Clean
the sample glasses before use as described in Chapter 7.2.3, page
36
.

2 Add nitric acid (w(HNO

3

) = 65%, for trace analysis) and hydrogen

peroxide (w(H

2

O

2

) = 30%, for trace analysis).

Samples with a low organic load are offset with the following
reagent concentrations:

0.01 mL HNO

3

(w(HNO

3

) = 65%, for trace analysis)

0.05 mL H

2

O

2

(w(H

2

O

2

) = 30%, for trace analysis)

The sample volume, the volume of the reagents and the digestion
parameters must be adjusted accordingly for samples with increased
organic load.

3 Place the PTFE stopper (6.1446.100) on the sample vessels. Leave the

sample vessel at Position 6 free for measuring the temperature.

4 Insert the sample holder into the wet end (see Chapter 3.5, page

17).

Note

Fully filled sample holder

Only a fully filled sample holder enables even temperature distribu-
tion to the samples. If you use fewer than twelve samples, fill the
remaining sample vessels with distilled water.

5 Insert the temperature sensor (6.1110.010) with attached thermom-

eter plug (6.1446.190) and spacer into the sample vessel at Position
6
and press it down.