Metrohm 797 VA Computrace User Manual
Page 93

4.1
Technical data
797 VA Computrace / Hardware-Manual 8.797.8001EN
85
Dummy Cell
Use
Checking of the 797 VA Computrace
Determination of the signal-to-noise ratio
Connections
AE
auxiliary electrode
RE
reference electrode
WE-L
working electrode linear mode
(RC element)
WE-D
working electrode differential mode
(peak/wave)
Inert gas (in general nitrogen N
2
)
Use
Operation of MME de-aeration of sample solution (Not
needed for electroplating bath analysis)
Purity
min. 5.0 (99.999 %)
Required pressure
p = 1 ± 0.2 bar (this gas pressure results in a gas flow
rate of ca. 20 L/h)
Dosing devices (Option)
Type
800 Dosino (2.800.0010) + 807 Dosing unit
(6.3032.XXX),
700 Dosino (2.700.0020) + 807 Dosing unit
(6.3032.XXX),
685 Dosimat (2.685.0010) + 806 Exchange unit
(6.3026.XXX) + Cable (6.2134.030)
805 Dosimat (2.805.0010) + 806 Exchange unit
(6.3026.XXX)
connect via MSB 1…3 (or to a 846 Dosing Interface)
Number
1...3 (with a 846 Dosing Interface, 4 additional can be
connected)
Plug
Mini-DIN
Manual operation
Dispensing, filling, adjustment of feed and filling rate
MSB connections
(MSB = Metrohm Serial Bus)
Dosing devices
Connection of max. 3 Dosing devices (MSB1 to MSB3)
to 797 VA Computrace. With a 846 Dosing Interface,
four additional Dosing devices can be connected.
USB connections
USB Standard
USB1.1
Connection computer
With cable 6.2151.020
USB ports
2 USB downstream Ports (type A sockets), each
500 mA, for the connection of peripherals as printer,…
Remote connections
Remote Interface
D-Sub, 25 pin remote interface to control sample
changer and rinsing equipment.