Rc sample with response curve, 2 installation, Measurement – Metrohm 797 VA Computrace User Manual
Page 82: Measure q(0), Measurement end

2 Installation
797 VA Computrace / Hardware-Manual 8.797.8001EN
74
2.13.6
Communication diagram "RC Sample with response curve"
797 VA Computrace
Dosinos 1..3
838 Advanced Sample
Processor
731 Relay Box
Set line: Start 838 (TTL 3)
Ready (TTL 7)
Set line: Siphoning (TTL 5)
Prep Dosino 1,2,3 ?
Prep Dosino 1, 2, 3
YES
Set line: Siphoning (TTL 5)
Wait siphoning time
Set line: Rinsing (TTL 6)
Wait rinsing time
No. of Prep
cycles
Is it the
first sample of the
sequence ?
NO
Set line: Siphoning (TTL 5)
Wait siphoning time
NO
YES
Set line: Transfer sample (TTL 4)
Ready (TTL 7)
Predose
Measurement
Set line: Siphoning (TTL 5)
Wait siphoning time
Set line: Rinse Capillary (TTL 4)
Ready (TTL 7)
Set line: Siphoning (TTL 5)
Wait siphoning time
Set line: Rinsing (TTL 6)
Wait rinsing time
No. of rinsing cycles
Export/Print results
Set line: End of sample (TTL 1)
More samples ?
NO
YES
Set line: End of sample table (TTL 2)
RC Sample with response curve
active ?
YES
NO
No. of rinsing cycles
Wait siphoning time
Set line: Siphoning (TTL 5)
Wait siphoning time
Measure Q(0)
Dose