4 operation of the 863 compact autosampler, Operation of the 863 compact autosampler – Metrohm 797 VA Computrace User Manual
Page 60

2 Installation
797 VA Computrace / Hardware-Manual 8.797.8001EN
52
2 Set automation parameters
• Select the
Automation
tab.
• Select „813/863 Compact Autosampler“ for the field Sample processor in
the Sample handling-part of the
Automation
tab.
• Check the checkbox
Relay box
.
• Normally, the default settings can be used for samples with 10 mL volume:
3 Test automation parameters
• Fill two sample vessels with water and place them in position 1 and 2 on
the sample rack of the 863 Compact Autosampler.
• Click on
797 VA COMPUTRACE
/ Settings / General settings
and select the
Auto-
mation
tab.
• Click on
, check the automation parameters and modify them if
needed.
2.9.4
Operation of the 863 Compact Autosampler
After installation of the instruments according to sections 2.9.1...2.9.3 sample series
using the 863 Compact Autosampler can be started. Proceed always in the following
sequence:
1 Switch on instruments
• Switch on PC.
• Switch on 797 VA Computrace Stand.
• Switch on 863 Compact Autosampler and 843 Pump Station.