Metrohm 797 VA Computrace User Manual
Page 50

2 Installation
797 VA Computrace / Hardware-Manual 8.797.8001EN
42
• Note: The maximum allowed dose rate is 2 mL/min if the 4-way microtip is
used to dose the solution in the measuring vessel.
• Click
• Click on
or
MAIN WINDOW / Utility / Dosino control
to open the
DOSINO
CONTROL
window.
• The connected Dosino and the mounted Dosing unit are automatically iden-
tified.
• Click the
button to empty and refill the Dosing unit installed on
the Dosino.
• Check if there are air bubbles left in the glass cylinder of the Dosing unit. If
this is the case, repeat the flushing procedure by clicking the
button.
• Close the
DOSINO CONTROL
window.
• Note: If you choose Port 3 for menu item
Prep / Empty via port
, you must
install an FEP Tubing Connection 6.1805.XXX from Port 3 to a waste con-
tainer.
685 Dosimat and 805 Dosimat
1 Mount Exchange unit on Dosimat
• Procedure, see Instructions for Use 685, section 2.1; and Instructions for
Use 805, section 3.1 respectively.