beautypg.com

Metrohm 847 Lab Link User Manual

Page 42

background image

4 Troubleshooting

Time for Periodic
status test
is set too
high

If the Lab Link Server list con-
tains a large number of non-
existent connections for which
the option Auto connect (via
multicast
) has been deactivated,
the ICMP timeout period must
be included when calculating
the time (see formula for Peri-
odic status test
, p. 13). We
recommend that unnecessary
connections are deleted from
the list.

Connection has been
interrupted with
Connection/ Discon-
nect

Turn off the Touch Control, wait
a few seconds, then switch it
back on.

Lab Link Server:

Status "not con-
nected"

Connection has only
been established
unidirectional

Restart the Lab Link Server.

Turn off the Touch Control, wait
a few seconds, then switch it
back on.

Touch Control Error
message:


"026-130 Incompati-
ble Software"

The software versions
of the Lab Link Server
and the Touch Con-
trols are not compati-
ble

On the Titrando installation CD
you will find versions that are
compatible.

38

847 USB Lab Link / Lab Link Server