2 Installation
2.3 847 USB Lab Link configuration
2.3.1 Connection to Titrando and network
Connect the 847 USB Lab Link to your Titrando-System as shown in Fig. 1
and connect it to the network. Proceed as follows:
1. Switch off your Titrando system.
2. Connect the USB connection 4 from the 847 USB Lab Link to a USB
connection on the Titrando. Use the 6.2151.030 USB cable.
3. Connect the RJ45 network connection 3 from the 847 USB Lab Link to
the local network using an RJ45 Ethernet cable.
Power supply to the 847 USB Lab Link is via the USB connection to the Ti-
trando.
Note
If free USB connections are no longer available on the Titrando you can use
a commercially available
USB hub. However, it is essential that this has its
own power supply, i.e. you must use a self-powered hub.
2.3.2 Parameter configuration
The following parameters are set directly on the Touch Control.
•
Device name for the 847 USB Lab Link
•
TCP/IP settings
•
Parameters for the PC/LIMS report
•
Parameters for messages
Note
Once these parameters have been set you must switch off the Touch
Control, wait a few seconds and then switch it on again so that the altera-
tions are adopted by the system.
1. Open the dialog System / Device manager.
2. Mark the entry 847 Lab Link and press
[Edit]
.
16
847 USB Lab Link / Lab Link Server