beautypg.com

Metrohm 847 Lab Link User Manual

Page 24

background image

2 Installation

More parameters

Under Edit device / TCP/IP setup you can use

[More Param.]

to access

the dialog in which the MAC address, manufacturer and product ID as well

as the device release are shown.

PC/LIMS Report

This is where you define where the machine-readable ASCII report is to be

stored. A PC/LIMS report contains all the data of a determination (method,

sample size, result, messages, etc.) and is saved as a *.txt file.

1. Touch the

[PC/LIMS Report]

field in order to make the settings for the

PC/LIMS report.

The checkbox Save PC/LIMS Report is always activated. You must define

the memory location for the report.

2. Touch the Memory field and select whether the report is to be stored on

Card 1, Card 2 or in the Shared memory.

Note
If you select the memory location (

Card 1 or Card 2) you must make sure

that a memory card is inserted in the corresponding slot.

20

847 USB Lab Link / Lab Link Server