6 control device, 7 sensor, Control device – Metrohm 840 PC Control 6.0 / Touch Control User Manual
Page 30: Sensor
![background image](/manuals/586607/30/background.png)
2.2 Bromine index determination (BRC)
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
Generator current
Polarization current at the generator electrode.
Selection
100 mA | 200 mA | 400 mA | auto
Default value
auto
auto
The current is adapted to the conductivity of the reagent and automat-
ically reduced near the endpoint.
Reagent monitoring
Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System
▶ Reagents.
Selection
Selection of configured reagents | off
Default value
off
off
The reagent monitoring is not active.
2.2.6
Control device
Under [Control device], the control device is selected the determination
is carried out with. Control devices are defined under System
▶ Device
manager.
Note
This button is only displayed when several control devices have been
configured.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.
Selection
Selection of configured control devices
2.2.7
Sensor
Under [Sensor], the parameters for the sensor are edited.
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System
▶ Sensors. You can
also enter a sensor name which is not contained in the sensor list. When a