2 titration parameters, Titration parameters – Metrohm 840 PC Control 6.0 / Touch Control User Manual
Page 17

■■■■■■■■■■■■■■■■■■■■■■
2 Parameters
■■■■■■■■
11
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.
off
The titration will not be stopped until the stop conditions have been
fulfilled.
Stop drift
This parameter can only be edited when Stop criterion = drift.
The titration is stopped when the endpoint and the stop drift have been
reached.
Range
1 ... 999 µg/min
Default value
5 µg/min
Delay time
This parameter can only be edited when Stop criterion = time.
When the endpoint has been reached, the specified time is allowed to
elapse after the last dosing and the titration is then stopped.
Range
0 ... 999 s
Default value
10 s
Relative stop drift
This parameter is can only be edited when Stop criterion = rel. drift.
The titration is stopped when the endpoint and the sum of the drift at the
start of the titration and the relative stop drift have been reached.
Range
1 ... 999 µg/min
Default value
5 µg/min
2.1.2
Titration parameters
Under [Titration parameters], the parameters influencing the sequence
of the entire titration are defined.
Pause
Waiting time, for example for dissolving the sample. During this time no
iodine is generated.
Range
0 ... 999999 s
Default value
0 s
Extraction time
Minimum duration of the titration. The titration will not be stopped during
the extraction time, even if the endpoint has already been reached. The
titration is however stopped if a stop condition is fulfilled during this time