Metrohm 840 PC Control 6.0 / Touch Control User Manual
Page 20
2.1 Coulometric Karl Fischer titrations (KFC)
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
Automatic start [Autom. start]
Automatic start
on | off (Default value: off)
If this parameter is activated, the determination is automatically started
when the measured voltage suddenly changes. This setting will be ignored
as long as the working medium has not been conditioned.
Threshold value
If the change in voltage is higher than this value, the determination will
automatically be started.
Range
0 ... 999 mV
Default value
50 mV
Conditioning options [Cond. options]
Stabilizing time
Waiting time, during which the measured drift has to be smaller than the
start drift defined until Conditioning OK is displayed. The start drift is
defined in the dialog Edit command / Conditioning.
Range
0 ... 999999 s
Default value
0 s
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Range
1 ... 999999 s
Selection
off
Default value
off
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.
Note
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.