Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 10

background image

1.1 Reagents

■■■■■■■■■■■■■■■■■■■■■■

4

■■■■■■■■

[Reagent exchange]

Edit the parameters for the reagent exchange.

Dialog "Reagent monitoring / Status"
In this dialog the current values of the reagent monitoring are displayed.

[Reset]

Reset the values to zero.

Dialog "Reagent monitoring / Reagent exchange"
In this dialog the parameters for the reagent exchange are defined.

Reagent exchange

The reagent can either be exchanged manually or automatically.

Selection

manual | auto

Default value

manual

manual
If a monitored parameter has reached the limit set, a message is being
displayed. Then the reagent has to be exchanged manually.
auto
If a monitored parameter has reached the limit set, the method defined
below is started automatically.

Memory

This parameter can only be edited when Reagent exchange = auto.

Memory location the method is loaded from. All memory locations are
selectable, even if they are currently not accessible.

Selection

Internal memory | Card 1 | Card 2 | Shared
memory

Default value

Internal memory

Shared memory
Memory released on an external system. For this an 847 USB Lab Link
is necessary.

Method

This parameter can only be edited when Reagent exchange = auto.

Method used for emptying the titration cell.

Note

Make sure that the memory is accessible.