2 Parameters

■■■■■■■■

19

Max. rate

Rate at which bromine is generated outside the control range.

Range

10.0 ... 19876.0 µg/min

Default value

1000.0 µg/min

Selection

maximum

Min. rate

Rate at which bromine is generated at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.

Range

1.0 ... 999.9 µg/min

Default value

25.0 µg/min

Table 2

Default values for the predefined titration rates for BRC

Titration rate

slow

optimal

fast

Dynamics

400.0 mV

400.0 mV

400.0 mV

Max. rate

500.0 µg/min

1000.0 µg/min

2000.0 µg/min

Min. rate

15.0 µg/min

25.0 µg/min

50.0 µg/min

Stop criterion

Stop criterion

The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 2.2.3,
page 21)
always lead to a stop, even if the stop criterion has not been
reached.

Selection

drift | time | rel. drift | off | drift & time

Default value

drift & time

drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped when the endpoint has been exceeded during
a certain period of time (Delay time).
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.