Metrohm 840 PC Control 6.0 / Touch Control User Manual
Page 25
■■■■■■■■■■■■■■■■■■■■■■
2 Parameters
■■■■■■■■
19
Max. rate
Rate at which bromine is generated outside the control range.
Range
10.0 ... 19876.0 µg/min
Default value
1000.0 µg/min
Selection
maximum
Min. rate
Rate at which bromine is generated at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.
Range
1.0 ... 999.9 µg/min
Default value
25.0 µg/min
Table 2
Default values for the predefined titration rates for BRC
Titration rate
slow
optimal
fast
Dynamics
400.0 mV
400.0 mV
400.0 mV
Max. rate
500.0 µg/min
1000.0 µg/min
2000.0 µg/min
Min. rate
15.0 µg/min
25.0 µg/min
50.0 µg/min
Stop criterion
Stop criterion
The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 2.2.3,
page 21) always lead to a stop, even if the stop criterion has not been
reached.
Selection
drift | time | rel. drift | off | drift & time
Default value
drift & time
drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped when the endpoint has been exceeded during
a certain period of time (Delay time).
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.