Metrohm 840 PC Control 6.0 / Touch Control User Manual
Page 16
2.1 Coulometric Karl Fischer titrations (KFC)
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
iodine is being continuously generated, the rate is defined under Max.
rate.
Range
0.1 ... 1250.0 mV
Default value
70.0 mV
Selection
off
Max. rate
Rate at which iodine is generated outside the control range.
Range
1.5 ... 2241.0 µg/min
Selection
maximum
Default value
maximum
Min. rate
Rate at which iodine is generated at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.
Range
0.3 ... 999.9 µg/min
Default value
15.0 µg/min
Table 1
Default values of the predefined titration rates for KFC
Titration rate
slow
optimal
fast
Dynamics
120.0 mV
70.0 mV
30.0 mV
Max. rate
1000.0 µg/min
maximum
maximum
Min. rate
0.3 µg/min
15.0 µg/min
30.0 µg/min
Stop criterion
Stop criterion
The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 2.1.3,
page 12) always lead to a stop, even if the stop criterion has not been
reached.
Selection
drift | time | rel. drift | off
Default value
rel. drift
drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped when the endpoint has been exceeded during
a certain period of time (Delay time).