2 Parameters
■■■■■■■■
23
Conditioning options [Cond. options]
Stabilizing time
Waiting time, during which the measured drift has to be smaller than the
start drift defined until Conditioning OK is displayed. The start drift is
defined in the dialog Edit command / Conditioning.
Range
0 ... 999999 s
Default value
0 s
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Range
1 ... 999999 s
Selection
off
Default value
off
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.
Note
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.
2.2.5
Cell
Under [Cell], the parameters for the titration cell are defined.
Generator electrode
Type of generator electrode.
Selection
without diaphragm | with diaphragm
Default value
with diaphragm
without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.