8 stirrer, Stirrer – Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 23

background image

■■■■■■■■■■■■■■■■■■■■■■

2 Parameters

■■■■■■■■

17

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature is meas-
ured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
will be used (see dialog of the titration and measuring parameters).
off
The temperature will not be measured. The temperature entered man-
ually is used (see dialog of the titration and measuring parameters).

2.1.8

Stirrer

Under [Stirrer], the parameters for the stirrer are edited.

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB
connector will always be displayed.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 3.1, page 27. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Range

–15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.