4.1.4 Mains
connection
Voltage
115 V: 100...120 V
±
10 %
230 V: 220...240 V
±
10 %
Switched with mains voltage selector in the fuse
Frequency
50...60 Hz
Power consumption
40 VA
Fuses
5 mm dia., 20 mm long
100
…
120 V: 1.0 A (slow blow)
220
…
240 V: 0.5 A (slow blow)
replace only with the same type
4.1.5 Safety
specifications
Construction and testing
According to EN/IEC 61010-1 / UL 3101-1,
protection class I
Safety information
The Instructions for Use contain safety informa-
tion that must be observed by the user in order to
ensure the safe operation of the instrument.
4.1.6
Electromagnetic compatibility (EMC)
Emission
Standards fulfilled:
- EN/IEC 61326-1
- EN 55022
Immunity
Standards fulfilled:
- EN/IEC 61326-1
- EN/IEC 61000-4-2
- EN/IEC 61000-4-3
- EN/IEC 61000-4-4
- EN/IEC 61000-4-5
- EN/IEC 61000-4-6
- EN/IEC 61000-4-11
- NAMUR
4.1.7 Ambient
temperature
Nominal working range
+5
…
+45 °C
(at 20
…
80% relative humidity)
Transport
–40
…
+70 °C
816 IC Eluent Selector/ 8.816.1003 Instructions for Use