2 fuses, 3 mains cable and mains connection, 4 switching the instrument on/off – Metrohm 816 IC Eluent Selector User Manual
Page 15: Fuses, Mains cable and mains connection, Switching the instrument on/off
2.2 Mains connection
2.2.2 Fuses
The fuse holder of the IC Eluent Selector is fitted with one of the follow-
ing two fuses as standard: 1.0 A/slow-blow for 100
…
120 V or
0.5 A/slow-blow for 220
…
240 V.
Never use different fuses in the instrument from those mentioned
above as otherwise there is a fire risk!
Checking and exchanging fuses is described in Section 2.2.1.
2.2.3
Mains cable and mains connection
Mains
The instrument is supplied with one of the following mains cables:
•
6.2122.020 with SEV 12 plug (Switzerland,
…
)
•
6.2122.040 with CEE(7), VII plug (Germany,
…
)
•
6.2133.070 with NEMA 5-15 plug (USA,
…
)
which has three wires and is fitted with a plug with a grounding pin. If a
different plug has to be used then the yellow/green wire (IEC standard)
must be connected to the grounding pin (protection class I).
Any interruption to the grounding inside or outside the instrument can
represent a hazard!
Mains connection
Insert the mains cable into mains connector 9 of the 816 IC Eluent Se-
lector.
2.2.4
Switching the instrument on/off
The 816 IC Eluent Selector is switched on and off with mains switch 8.
816 IC Eluent Selector/ 8.816.1003 Instructions for Use
9