Figure 3, Setting the mains voltage – Metrohm 816 IC Eluent Selector User Manual
Page 14
2 Installation
1 Pull out mains cable
Remove the mains cable from mains supply connection 9 of the
816 IC Eluent Selector.
2 Remove the fuse holder
Use a screwdriver to loosen fuse holder 12 below the mains
supply connection and remove it completely.
3 Check the fuse
Carefully remove the built-in fuse for the intended voltage from the
fuse holder and check its specifications (the position of the fuse in
the fuse holder is indicated by the white arrow beside the voltage
range):
100
…
120 V 1.0 A (slow blow) Metrohm No. U.600.0016
220
…
240 V 0.5 A (slow blow) Metrohm No. U.600.0013
4 Insert fuse
Exchange the fuse if necessary and replace it in the fuse holder.
5 Insert fuse holder
Depending on the required mains voltage, insert the fuse holder
in the 816 IC Eluent Selector so that the correct voltage range can
be read normally and the white arrow beside it points to the white
bar printed below the fuse holder (see Figure 3).
220 – 240 V
100 – 120 V
8 Mains switch
9 Mains connec-
tion
12 Fuse holder
220
- 240 V
100 - 1
20
V
220 - 2
40
V
100
- 120 V
12
9
8
Figure 3:
Setting the mains voltage
816 IC Eluent Selector/ 8.816.1003 Instructions for Use