Metrohm 827 pH lab User Manual
Page 54

8.2 Messages
46
826/827 pH meter, Instructions for Use
Message
Situation
Cause
Remedy
battery low
Mode pH/U/T
The batteries are almost
empty
Change batteries (see Sec-
tion 8.4.1)
buffer error
calibration
Buffer not recognized
1. Replace buffer
2. Press
buffer not de-
fined
calibration
pH value of a buffer has
not been defined for the
given temperature
pH values are only defined
in the range of 0 ... 95 °C
calibration OK calibration
Calibration has been con-
cluded successfully
Exit display with
tomatically after 30 s; cali-
bration data will be ac-
cepted in either case
cal. interval
exp.
mode pH
Set calibration interval has
expired
Recalibrate electrode (see
Section 6.1.2
cal. out of
limits
calibration
Determined calibration
data are outside limits de-
fined as calibration para-
meters
Accept calibration data with
buffer or limits under 6.1.2 /
6.1.3
change buffer
calibration
Instrument is waiting for
new buffer
1. Change buffer
2. Press
check T sensor
Mode T
Temperature sensor is faul-
ty or not connected
Connect a properly function-
ing temperature sensor
delta T >2 °C calibration
Temperature difference
between two buffers is too
large
1. Match temperatures
2. Continue with
cancel with
<
QUIT> and
restart
Err #X
System test after
switch on
pH meter fault
1. Note error number
2. Contact Metrohm Service
IR off
Print report /
measured value
The IR interface is not acti-
vated
Activate the IR interface (see
Section 5.4
manual stop
calibration
The calibration has been
stopped manually
mv memory
overflow
Saving measured
values
Attempt made to store a
further measured value
when 200 have already
been stored
Delete measured values
(see
Section 7.2
no IR connec-
tion
No connection with printer 1. Press
2. Switch on printer
3. Switch pH meter off and
on again