5.2 Printing out measured values
24
826/827 pH meter, Instructions for Use
adapted accordingly, as otherwise the buffer memory of the con-
nected printer would be filled too quickly. This means, for exam-
ple, that the report header should only be printed out once at the
start of a series of measurements. Otherwise data could be lost.
In this case you should consider matching the measured value
recording to the measured value memory (see Section 5.3).
stop time:
0...999'999 s
The time after which an automatic series of measurements is to
be terminated can be entered in whole seconds. You can also let
a series of measurements run indefinitely without a fixed stop time
and then stop it manually with
value'0 s' for this setting.
print crit.:
drift
If this print criterion is used, a measured value will be printed out after
pressing
ing... is no longer displayed.
date & time:
on, off
rep.header:
once, always, off
cal.report:
on, off
These settings correspond to those of the print criterion 'immed.'
(see above).
print crit.:
change
The next measured value is printed automatically not until the difference
to the previous measured value is bigger than the parameter delta
pH/T/mV defined here.
date & time:
on, off
rep.header:
once, always, off
cal.report:
on, off
These settings correspond to those of the print criterion 'immed.'
(see above).
delta pH:
0.10...0.50...16.00 pH
delta T:
0.1...0.5...100.0 °C
delta mV:
0.1...30.0...999.9 mV
stop time:
0...999'999 s
The printout of the measured values is terminated when the stop
time has elapsed or will continue indefinitely ('0 s'). A termination
is always possible with