8.1 Troubleshooting
44
826/827 pH meter, Instructions for Use
Problem
Possible cause
Remedy
Measuring signal absent
or very variable
Electrode not connected
•
Connect electrode
Air in or in front of the diaph-
ragm
•
Remove air
Faulty electrode
•
Replace electrode
Measured value remains
unstable and does not
fulfill drift criterion
Dirty glass membrane or diaph-
ragm
•
Clean membrane or diaphragm
pH or temperature of the solu-
tion is not stable
•
Measure under exclusion of air
•
Thermostat the solution
Unsuitable electrode:
•
Conductivity too low
•
Organic solution
•
Use suitable electrode
Electrode not connected
•
Connect electrode
Faulty electrode
•
Replace electrode
Slow adjustment of
measured value
Dirty glass membrane or diaph-
ragm
•
Clean membrane or diaphragm
Slope too small on cali-
bration
Dirty glass membrane or diaph-
ragm
•
Clean membrane or diaphragm
Glass membrane deswollen af-
ter measurement in non-
aqueous solution
•
Immerse electrode in water between
measurements
Poor buffer solutions
•
Replace buffers
Old Electrode
•
Replace electrode
Measured value obvious-
ly incorrect
Incorrect calibration
•
Check / Repeat calibration
•
Check / Replace buffers
Manual temperature input incor-
rect
•
Enter correct temperature of the solu-
tion
Dirty glass membrane or diaph-
ragm
•
Clean membrane or diaphragm
Electrolyte or electrode too old
•
Refill electrolyte or replace electrode
Faulty electrode
•
Replace electrode