6 Methods / Parameters

826/827 pH meter, Instructions for Use

31

no.of buffers

1...2...3

Up to 3 buffers can be used for the pH calibration. If the calibration is

carried out with only one buffer then the program will use the theoretical

value of 100.0 % for the slope. This also applies when a calibration with

several buffers is canceled with after the first buffer has been

measured and then confirmed with .
Enter "3" if you carry out a 2-point calibration but wish to measure the

second buffer twice in order to give it more statistical weight.

buff.type:

Metrohm, NIST, DIN, Fisher,
Fluka BS, Mettler, MerckTitrisol,
MerckCertiPUR, Beckman, Radiome-
ter, Baker, Hamilton, Precisa,
special

For the automatic temperature-specific buffer recognition during the ca-

libration, information about the type of buffer used is required. In the pH

meter the temperature-dependent pH values of reference buffer solu-

tions and technical buffer solutions from some suppliers are stored.

Such a buffer table for Metrohm buffers looks like this:

Metrohm buffers

T [°C] pH 4.00 pH 7.00 pH 9.00

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

3.99

3.99

3.99

3.99

3.99

4.00

4.00

4.01

4.02

4.03

4.04

4.06

4.07

4.09

4.11

4.13

4.15

4.18

4.20

4.23

7.11

7.08

7.06

7.04

7.02

7.00

6.99

6.98

6.98

6.97

6.97

6.97

6.97

6.98

6.98

6.99

7.00

7.00

7.01

7.02

9.27

9.18

9.13

9.08

9.04

9.00

8.96

8.93

8.90

8.87

8.84

8.81

8.79

8.76

8.74

8.73

8.71

8.70

8.68

8.67