4 short operating tutorial, 1 requirements, 2 ph calibration – Metrohm 827 pH lab User Manual
Page 25: Short operating tutorial, Requirements, Ph calibration, 4short operating tutorial

4 Short operating tutorial
826/827 pH meter, Instructions for Use
17
4
Short operating tutorial
In this section the necessary steps for carrying out a simple pH and ion
measurement with calibration are described. The instructions are li-
mited to those steps that are absolutely necessary and will enable you
to carry out your first measurements with the pH meter directly. The op-
erating principles are described in Section 3.3.
4.1
Requirements
The following instruments, accessories and solutions are required for
carrying out the pH calibration and measurement described below:
•
826 pH mobile (2.826.0XX0) or 827 pH lab (2.827.011X,
2.827.021X)
•
pH electrode
•
Calibration buffers
Metrohm buffer solutions pH 4.00 and pH 7.00
The calibration parameters for the pH mode are set for a calibration us-
ing two Metrohm buffers as default (see Section 6.1.2). If you want to
use other buffers then the corresponding buffer type must be entered.
4.2
pH calibration
1 Start calibration with first buffer
•
Start calibration with
•
Immerse pH electrode in buffer solution pH 7 and confirm with
•
If a temperature sensor is connected, the calibration tempera-
ture is being determined
•
If no temperature sensor is connected:
enter the temperature with the arrow keys and confirm with
<OK>
•
The first buffer is being measured
•
The following message is displayed after successful mea-
surement:
change buffer
2 Continue calibration with second buffer
•
Remove pH electrode from first buffer and rinse with water
•
Immerse pH electrode in second buffer solution pH 4 and
continue calibration sequence with
•
The second buffer is being measured