7 Various functions
826/827 pH meter, Instructions for Use
37
7.1.3
Measuring point report
By pressing the
ing point report will be produced. The following examples show ver-
sions of this type of report, the various configurations of which are de-
fined under configuration/print (see Section 5.2).
A typical version of a measuring point report is the output of a single
measured value. The print criterion 'immed.' or 'drift' must be selected.
'mp
827 pH lab
01106 5.827.0010
print date
2005-01-27 14:24:38
pH = 6.958
23.6 °C NTC
==============
If a whole series of measured values is to be documented then this is
done by selecting the option rep.header: once under configura-
tion/print. In this way the report header will be printed out once with
the first measured value.
Note!
In order that the report header is printed again for a new series of
measured values, the pH meter must be switched off and on again.
Each further time that the
value will be printed out:
pH = 6.963
23.6 °C NTC
pH = 6.968
23.6 °C NTC
pH = 6.969
23.6 °C NTC
Such a series of measured values can also be produced automatically
by selecting the print criterion 'time'.
With the print criterion 'change' the next measured value is printed au-
tomatically not until the difference to the previous measured value is
bigger than the value delta pH/T/mV defined under configura-
tion/print.
7.1.4
Calibration report
Subsequent to a calibration a calibration report can printed out. Under
param/cal.settings/report you can define whether this report is to
be produced automatically after the measurement.
7.1.5
Configuration report
The current device settings are printed out with a configuration report.
<< print >>
<< report >>
cal
<< report >>
config