5 Configuration
826/827 pH meter, Instructions for Use
25
5.3
Storing measured values
<< config >>
configuration
:
├
store
: └
store crit.
┬
immed., drift, off
├
time
┬
time int.
│
└
stop time
├
change
┬
delta pH/T/mV
└
└
stop time
Up to 200 measured values can be stored manually or automatically af-
ter reaching the drift criterion, at definite time intervals or after changing
by a certain amount (see Section 7.2). The storing is done by pressing
store crit.:
immed., time, drift, change, off
Depending upon the store criterion a measured value will be stored ei-
ther immediately, at defined time intervals, when the drift condition
has been fulfilled or after a certain difference of the measured value
has been exceeded.
The following additional information are stored with each measured val-
ue:
•
Temperature (Mode pH only)
•
Id
•
Date
•
Time
store crit.:
immed.
The current measured value is stored with
menu << store >>.
store crit.:
time
time int.:
1...4...99'999 s
stop time:
0...999'999 s
You can automatically record and save the measured values at fixed
time intervals until the stop time has been reached.
store crit.:
drift
A measured value can only be stored if the message drifting... is
no longer displayed.
store crit.:
change
The next measured value is stored automatically not until the difference
to the previous measured value is bigger than the parameter delta
pH/T/mV defined here.