4.3 Technical Specifications
760 Sample Changer, Appendix
29
4.3 Technical Specifications
Dimensions
Height: 0.79 m, Width: 0.28 m, Depth: 0.48 m
Weight
12.5 kg
Material
Sample Changer case:
metal case, multiple enamelling
Splash protection:
Polymethylene methacrylate (PMMA)
Sample rack:
Acrylnitrile butadiene styrene (ABS)
or Polypropylene (PP)
Lift path
235 mm
Lift
Load:
approx. 10 N
Stroke speed: adjustable, 3...25 mm/s
Turntable
Turning speed: 20 angular degrees/s
Stirrer
Stirring rate: adjustable to 15 levels
- magnet stirrer 180/min...2600/min
- rod stirrer
180/min...3000/min
Remote Interface Programmable parallel interface for controlling external
instruments
Input:
t
p
t
p
>20 ms
Output:
t
p
t
p
>200 ms
V
CEO
= 40 V
I
C
= 20 mA
+5 V:
maximum load = 20 mA
Temperatures
Nominal functional range 5...40 °C
at 20…80% rel. humidity
Transport und storage –20...+60 °C
60 °C rel. humidity < 50%
50 °C rel. humidity < 85%
40 °C rel. humidity <95%
Power connection Voltage
100...120 V, 220...240 V
Frequency
50...60 Hz
Power input
40 VA
Fuses
0.5 AT (110 V), 0.25 AT (220 V)
All data are typical values with the exception of those specially marked.
Safety Specifications
Constructed and tested according to IEC 1010 / EN 61010 /
UL 3101-1, Safety Class III
Degree of protection IP 22