beautypg.com

Metrohm 760 Sample Changer User Manual

Page 27

background image

4.1 Error Messages

760 Sample Changer, Appendix

24

9* missing beaker

After a MOVE command no beaker could be found at
the selected position.

10* raise lift first

Turning of a rack could not be carried out because a
lift was below the defined shift position.

12* changer overload

Load or resistance too large to carry out the chosen
action.

14* changer not ready

The changer cannot execute the command chosen
because it is busy carrying out another action.