2.4 Remote Connections

760 Sample Changer, Installation

12

14 lines (Output 0–13) are available for the emission of signals.
For receiving signals 8 lines are provided (Input 0–7). The
assignment of these lines is fixed for the 760 Sample Changer.

Pin assignment of the Remote socket:

Outpu t 6
Outpu t 7
Outpu t 4
Outpu t 2
Outpu t 0
Outpu t 8
Outpu t 9
Outpu t 10
Input 1
Input 3
Input 5
Input 7
Outpu t 11

0 Vo lt

+5 Volt

Output 5
Output 3
Output 1

Output 12
Output 13

Input 0
Input 2
Input 4
Input 6

0 Vo lt



The +5 V supply line may by charged with 20 mA maximally.

2.4.1 Typical Combinations of Instruments

760 Sample Changer — Titrino

with standard cable

B

A

Titrino

C

D



Remote control commands:

Output 0:

*************1

starts Titrino

Input 3:

****1***

awaits end of titration (EOD pulse)

Cable 6.2141.020

Inputs

Outputs

+5V

active = low

inactive = high

active = low