2.4 Device connection at the MSB
14
Titrando Installation Instructions
plifiers can be connected in series; this allows a maximum extension of
25 m.
Start the PC Control or tiamo software.
The Titrando will be recognized automatically. When the PC Control or
tiamo software is started a system test will be carried out automatically
on the Titrando. The LED "On" on the Titrando lights up when the sys-
tem test is finished and the instrument is ready for use.
2.4 Device connection at the MSB
The following devices can be connected via each of the four MSB con-
nections (Metrohm Serial Bus): one stirrer or titration stand, one Do-
simat or Dosino dosing device and one remote box. The stirrers and
remote boxes each have an MSB output so that the devices can be
switched sequentially ("daisy chain"). The following illustration provides
an overview of the devices that can be connected to an MSB.
Fig. 8: Overview of MSB connections
The MSB 1 of the Titrando with internal dosing drive is occupied by the
internal dosing drive. This means that only a stirrer and a remote box
can be connected to MSB 1.
First connect all peripheral devices and then connect the Titrando to the
mains supply. MSB connections can be extended with the 6.2151.010
cable. The maximum length of the connection is 15 m.
The Titrando automatically recognizes which device has been con-
nected to which MSB connection. The Touch Control or the PC software
(PC Control or tiamo) shows the connected devices in the device man-
ager or the configuration dialog respectively. All devices connected to
the MSB are operated by the Touch Control or PC Control/tiamo.