2 instrument setup, 1 packaging, 2 checks – Metrohm 835 Titrando User Manual
Page 17: 3 location, 3 controller connection, Instrument setup, Packaging, Checks, Location, Controller connection
![background image](/manuals/586306/17/background.png)
2
Installation
Titrando Installation Instructions
11
2.2 Instrument
setup
2.2.1 Packaging
The Titrando and the separately packed accessories are delivered in
special packaging that provides excellent protection. Please store this
packaging in a safe place as no other packaging can guarantee the
safe transport of the instrument.
2.2.2 Checks
Please check whether the delivery is complete and undamaged imme-
diately on receipt (compare with delivery note and list of accessories
given in Section 4.2). If transport damage is evident please refer to the
information given in Section 4.4.1.
2.2.3 Location
The Titrando was developed for internal laboratory use; it should not be
used in explosion-endangered locations.
Place the instrument on a suitable vibration-free laboratory bench, pro-
tected as much as possible from corrosive atmospheres and contact
with chemicals.
Choose a location where the temperature is usually between +5 °C and
+45 °C. The instrument should be protected against excessive varia-
tions in temperature and direct sunlight.
2.3 Controller
connection
Two different methods of controlling the Titrando are available:
• The Touch Control with contact-sensitive screen forms a “stand-
alone” titrator together with the Titrando.
• A computer can be used to control the Titrando by using the PC
software PC Control or
tiamo.
Attention!
Make sure that the mains cable has been removed from the power
supply socket before setting up or breaking connections between the
instruments.