Maintenance

15

y

Turn off the RF power and interlock circuitry.

y

Disconnect the RF line.

y

Connect the multimeter test leads to the center and outer
conductor of the load. Refer to Figure 6.

y

Compare the measured value with the previous measurement and
with the baseline resistance, measured when the load was put
into service. If the new value differs from either of these by more
than 2 ohms this could indicate a failing resistor.

Coolant

Changing Coolant

Follow these instructions to change the coolant. To just add coolant,
go to step 5. To just drain the coolant, follow steps 1 – 4.

Figure 6

Measuring

Resistance

WARNING

Never attempt to connect or disconnect RF equipment from the

transmission line while RF power is being applied.

Leaking RF energy is a potential health hazard.

WARNING

Ethylene glycol is toxic. Do not take internally. Avoid contact with

eyes, skin, and clothing. Avoid breathing vapor. Wash thoroughly

after handling.

CAUTION

Use only distilled water or the supplied ethylene glycol as coolant.

Do not use tap water, automotive antifreeze, sealants, or leak

stopping material. Use of these materials will damage the

instrument and void all warranties.

CAUTION

Operation without sufficient coolant can damage the unit.