Normal Operation

y

Check that the coolant level is above the min. mark on the gauge.

y

Turn on the Moduload.

y

Check that the fans are running properly.

y

Wait about 2 seconds for proper coolant flow and for the interlock
to close.

y

Apply RF power.

Shutdown

y

Turn off RF power at the source.

y

Wait five minutes for the system to cool to room temperature.

y

Turn off the Moduload.

Emergency

Shutdown

Turn off RF power at the source.

If the interlock is properly connected, RF power will be automatically
turned off if a fault occurs in the Moduload.

CAUTION

Operation without sufficient coolant can damage the unit.

CAUTION

Do not operate without the interlock. Even momentary

application of RF power while coolant is not circulating could

cause immediate destruction of the load.